Measurement range
|
0 % ... 100 % oxygen (at atmospheric pressure) |
Nominal sensor lifetime
|
≥ 1 000 000 % volume oxygen hours |
Output in ambient air
|
13 mV ... 16 mV
|
Electrical interface
|
3.5 mm mono phone Jack
|
Accuracy
|
meets ISO 80601-2-55 requirements
|
Repeatability
|
< 1 % volume O2 at constant temperature and pressure
|
Linearity error
|
< 3 % relative
|
Response time
|
< 12 s to 90 % of final value
|
Zero offset voltage
|
< 200 µV in 100 % nitrogen, applied for 5 min
|
Cross interference
|
meets ISO 80601-2-55 requirements |
Influence of humidity
|
-0.03 % rel. per % RH at 25 °C |
Pressure range | 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2) |
Influence of pressure | proportional to change in oxygen partial pressure |
Influence of mechanical shock | < 1 % relative after a fall from 1 m |
Operating temperature | 0 °C ... +50 °C |
Temperature compensation
|
built-in NTC compensation |
Effect of temperature compensation (steady
state)
|
between +25 °C and +40 °C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
|
Operating humidity
|
0 % ... 99 % RH non-condensing |
Long term output drift
|
1 % volume oxygen per month
typically < -15 % relative over lifetime
|
Storage temperature
|
-20 °C ... +50 °C |
Recommended storage
|
+5 °C … +15 °C |
Recommended load
|
≥ 10 kOhms |
Warm-up time | < 30 minutes, after replacement of sensor |
Weight | approximately 28 grams |
1. In-Warranty Period: we will give the replacement for all quality problems in time under normal conditions of use.
2. If the new items can not work with your devices, we will send new replacement or refund in time
Brand Name: | original |
Model Number: | OOM202-1 |
MOQ: | 1pcs |
Price: | as demand |
Packaging Details: | 1pcs / bag |
Payment Terms: | T/T, Western Union, |
Plastic Medical Oxygen Sensor 3 Pin Envitec O2 Sensor 16mV Signal Output
Measurement range
|
0 % ... 100 % oxygen (at atmospheric pressure) |
Nominal sensor lifetime
|
≥ 1 000 000 % volume oxygen hours |
Output in ambient air
|
13 mV ... 16 mV
|
Electrical interface
|
3.5 mm mono phone Jack
|
Accuracy
|
meets ISO 80601-2-55 requirements
|
Repeatability
|
< 1 % volume O2 at constant temperature and pressure
|
Linearity error
|
< 3 % relative
|
Response time
|
< 12 s to 90 % of final value
|
Zero offset voltage
|
< 200 µV in 100 % nitrogen, applied for 5 min
|
Cross interference
|
meets ISO 80601-2-55 requirements |
Influence of humidity
|
-0.03 % rel. per % RH at 25 °C |
Pressure range | 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2) |
Influence of pressure | proportional to change in oxygen partial pressure |
Influence of mechanical shock | < 1 % relative after a fall from 1 m |
Operating temperature | 0 °C ... +50 °C |
Temperature compensation
|
built-in NTC compensation |
Effect of temperature compensation (steady
state)
|
between +25 °C and +40 °C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
|
Operating humidity
|
0 % ... 99 % RH non-condensing |
Long term output drift
|
1 % volume oxygen per month
typically < -15 % relative over lifetime
|
Storage temperature
|
-20 °C ... +50 °C |
Recommended storage
|
+5 °C … +15 °C |
Recommended load
|
≥ 10 kOhms |
Warm-up time | < 30 minutes, after replacement of sensor |
Weight | approximately 28 grams |
1. In-Warranty Period: we will give the replacement for all quality problems in time under normal conditions of use.